Omron NE- C801KD instruction manual Disinfecting the nebulising parts, Handling the Vaporiser Head

Models: NE- C801KD

1 26
Download 26 pages 50.75 Kb
Page 17
Image 17
4.2Disinfecting the nebulising parts

4.Cleaning and daily disinfecting

4.2Disinfecting the nebulising parts

Always clean before disinfection. (See Section 4.1 “Cleaning”)

Always disinfect the medication tank, mouthpiece and other nebulising parts after the last treatment of the day. For appropriate method of disinfecting each part, refer to page 18.

To disinfect the parts, use one of the methods described below:

A. Use a commercially available disinfectant. Follow the instructions provided by the disinfectant manufacturer.

1.Submerge the parts in the cleansing solution for the specified period.

2.Remove the parts and discard the solution.

3.Rinse the parts with clean hot tap water, shake off excess water and allow to air dry in a clean environment.

B.Parts may be boiled between 15 to 20 minutes.

After boiling, carefully remove the parts, shake off excess water and allow to air dry in a clean environment.

Note: Do not boil air tube, mask, kids accessory, air filter and air filter cover.

Handling the Vaporiser Head Caution:

Do not use an autoclave, EOG gas disinfection or low temperature plasma sterilizer to disinfect the device.

Note: The printing may come off the surfaces of kids accessories as a result of repeated washing and disinfecting.

Handling the Vaporiser Head

Baffle

 

The vaporiser head is an important part that is used to vaporise medication.

Be sure to observe the following precautions when handling it.

Manual background Caution:

• Always wash the vaporiser head after each use.

Do not use a brush or pin etc., to clean the parts.

When disinfecting the parts by boiling, be sure to boil them in plenty of water.

• Do not boil vaporiser head together with other objects, other than

Vaporiser Head

applicable nebuliser accessories.

 

• Do not use a microwave oven, dish dryer or hair dryer to dry parts.

EN

17

Page 17
Image 17