7. Troubleshooting
Check the following if your unit should fail during operation. You can also refer to the pages of this manual for complete instructions.
Problem | Cause | Remedy | |
|
|
| |
Nothing happens when the | Is the AC adapter connected | Check that the plug is | |
power switch is pressed. | correctly to the an electrical | inserted in an electrical | |
| outlet and the compressor? | outlet. Unplug then reinsert | |
|
| the plug if necessary. | |
No nebulisation or low | Is there medication in the | Add the correct amount of | |
nebulisation rate, when the | medication tank? | medication to the medication | |
power is on. |
| tank. | |
Is there too much/little | |||
|
| ||
| medication in the medication |
| |
| tank? |
| |
| Is the vaporiser head missing | Attach the vaporiser head | |
| and not assembled correctly? | correctly. | |
| Is the nebuliser kit | Assemble the nebuliser kit | |
| assembled correctly? | correctly. | |
| Is the nozzle blocked? | Make sure that the nozzle is | |
|
| free of blockages. | |
| Is the nebuliser kit tilted at a | Make sure that the nebuliser | |
| sharp angle? | kit is not tilted at an angle of | |
|
| more than 45 degrees. | |
| Is the air tube connected | Make sure that the air tube is | |
| correctly? | correctly connected to the | |
|
| compressor and nebuliser kit. | |
| Is the air tube folded or | Make sure that the air tube | |
| damaged? | does not contain kinks. | |
| Is the air tube blocked? | Make sure that the air tube is | |
|
| free of blockages. | |
| Is the air filter dirty? | Replace the air filter with a | |
|
| new one. | |
The device is abnormally | Is the air filter cover attached | Attach the air filter cover | |
loud. | correctly? | correctly. | |
The device is very hot. | Is the compressor covered? | Do not cover the compressor | |
|
| with any type of cover during | |
|
| use. | |
| Is the device operating | Limit use to 20 minutes at a | |
| continuously over 20 | time, and allow a 40 minute | |
| minutes? | interval before using the | |
|
| device again. |
Note: If the suggested remedy does not solve the problem, do not try to repair the device - no parts of the unit are user serviceable.
Return the unit to an authorized OMRON retail outlet or distributor.