Adjustment
1Center the plate vial bubble.
2Rotate the instrument 180°, move the bubble half way back to the center by operating a leveling screw which is located parallel with the plate vial.
3Rotate the vial adjusting screw with the adjusting pin and position the bubble in the center.
4Repeat procedure again to confirm.
38