27
Philips Medical Systems
APPENDICES
C Glossary of symbols/controls
symbol description
On/Off button. Green. Pressing the On/Off but ton when the defibrillator is in
standby mode turns the defibrillator on; pressing a nd holding the On/Off button
for one second when the defibrillator is on turns t he defibrillator off and disarms
the defibrillator. In addition, pressing the On/Off butt on stops the battery
insertion self-test that automatically runs when a battery is inserted.
Information button (i-button). Blue. Pressing the i-button while it is flashing
during a patient care pause provides CPR Coaching in default configuration;
pressing it while it is flashing and the defibrillator is chirping provides
troubleshooting guidance. Pressing it until it beeps at other times provides
summary information about the defibrillato r’s last clinical use. Pressing it briefly in
standby mode gives device status.
Caution light. Flashes during rhythm analysis, and is on but not flashing when a
shock is advised, as a reminder not to touch the patient.
Shock button. Orange. If a shock is nee ded, flashes when the defibrillator is
charged. The defibrillator directs the user to press the Shock button to deliver a
shock to the patient.
Refer to operating instructions.
Lithium manganese dioxide battery.
TSO C-142 certified battery (989803139301 only)
One battery in package.
Do not crush the battery.