39
Philips Medical Systems
APPENDICES

M5070A battery and 989803139301 TSO C-142* battery

Infant/Child Key 989803139311
* The conditions and tests required for TSO approval of this article are minimum performance standards. It is the responsibility of those
installing this article either on or within a specific type or class of aircraft to determine that the aircraft installation conditions are within
the TSO standards. TSO articles must have a separate approval for installation in an aircraft. The article may be installed only if performed
under 14 CFR Part 43 or the applicable airworthiness requirements.
category specifications
battery type 9 VDC, 4.2 Ah, lithium manganese dioxide. Disposable, long-life primary cell.
capacity When new, a minimum of 200 shocks or 4 hours of operating time at 25° C
(77° F). (IEC 60601-2-4, 2002)
shelf life
(prior to insertion)
A minimum of 5 years from date of manufacture when stored and maintained
according to directions provided in this document.
standby life
(after insertion)
Typically, 4 years when stored and maintained according to directions provided
in this document.
training life Supports 10 hours of use in training mode.
battery limitations Never charge, short circuit, puncture, deform, incinerate, heat above 60° C, or
expose contents to water. Remove the battery when discharged.
maintenance and calibration
requirements for continued
airworthiness
(989803139301 only)
There are no periodic maintenance or calibration requirements that are
necessary for continued airworthiness of the 989803139301 battery. For
battery maintenance with respect to performance within the AED, please see
Chapter 5. There are no user-serviceable parts in the battery.
environmental qualification
per RTCA/DO-227, Section
2.3
Meets following acceptance criteria: No leaking, venting, distortion, fire, or
rupture. Change in open circuit voltage <2%.
category specifications
size 6.3” x 2.4” x 0.2” (16 cm x 6 cm x 0.5 cm).
weight 1.0 oz (29 g).
material Polycarbonate.