38
Philips Medical Systems
HEARTSTART FRx 861304 OWNER’S MANUAL
APPENDICES
Accessories specificatio ns

HeartStart SMART Pads II 989803139261

non-shockable rhythm — all
other non-shockable
rhythmsc
450 specificity >95% (88%)
a. From Philips Medical Systems Heartstream ECG rhythm databases.
b. American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy. Automatic External Defibrillators for Public
Access Use: Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporation of New
Waveforms, and Enhancing Safety. Circulation 1997;95:1677-1682.
c. Supraventricular tachycardia (SVT) is specifically included in the non-shockable rhythm class, in accordance with AHA recommendationsb
and the AAMI standard DF80.
rhythm class
ECG test
samplea size
meets AHA recommendationsb for adult defibrillation
observed 90% one-sided
performance lower
confidence limit
category specifications
pads for defibrillation,
pacing, monitoring,
cardioversion
Disposable, adhesive pads with a nominal active surface area of 80 cm2 each,
provided in a disposable plastic case, and an integrated 48 inch (121.9 cm),
typical, cable. Pads in case are designe d to fit into carry cases.
SMART Pads II compatibility defibrillator model adult
patient use
infant/child patient use
FRx*
FR2/FR2+
FR/ForeRunner
MRx/XL/XLT/4000
HS1/OnSite/Home
competitive adapters
yes
yes
yes
yes
no; use M5071A
yes
yes
no, use M3870A
no
manual mode only
no; use M5072A
manual mode only
* Pre-connectable to FRx defibrillator only.
pads shelf life Pads package is labeled with a use-by date of at least two years from date of
manufacture.