Philips Medical Systems
HEARTSTART FRx 861304 OWNER’S MANUAL
APPENDICES
energy Using HeartStart SMART Pads II for adult defibrillation: 150 J nominal (±15%)
into a 50 ohm load.
Using HeartStart SMART Pads II with Infant/ Child Key inserted: 50 J nominal
(±15%) into a 50 ohm load. Sample pediatric energy doses:
age energy dose
newborn 14 J/kg
1 year 5 J/kg
2 - 3 years 4 J/kg
4 - 5 years 3 J/kg
6 - 8 years 2 J/kg
Doses indicated are based on CDC growth char ts for the 50th percentile
weights for boys.*
* National Center for Health Statistics in collaboration with the National Center for
Chronic Disease Prevention and Health Promotion. CDC growth charts: weight-for-age
percentiles, revised and corrected November 28, 2000. Atlanta, GA: Centers for Disease
Control and Prevention © 2000.
charge control Controlled by Patient Analysis System for automated operation.
shock-to-shock cycle time < 20 seconds typical, in cluding analysis.
“charge complete”
indicator
Shock button flashes, audio tone sounds; device is able to deliver a shock as
soon as a shock is advised.
patient care
pause-to-shock time
Quick Shock. 8 seconds, typical , from end of patient care pause to shock
delivery.
disarm (AED mode) Once charged, the HeartStart FR x will disarm if:
patient’s heart rhythm changes to non-shockable rhythm,
a shock is not delivered within 30 seconds after the FRx is armed,
the On/Off button is pressed for one second to turn off the FRx,
the Infant/Child Key is inserted or removed,
the battery has been removed or is completely depleted, or
the impedance between pads is out of range.
adult shock delivery vector Via SMART Pads II placed in the anterio r-anterior (Lead II) position.
infant/child shock delivery
vector
Via SMART Pads II typically placed in the anterior-posterior position.
category nominal specifications