Pressure Systems, Inc. | 9016 Upgrade Instructions |
Remove 12 perimeter Phillips- head screws.
REMOVE ONLY
THE 12
PERIMETER
SCREWS.
Figure 3
9016 Top Plate
(2)When all screws have been removed, gently lift the top panel and attached electronics up and out of the housing. All components of the pressure scanner are attached to the top plate and will lift out of the module housing when the top plate is removed. See Figure 4, next page. In some cases, it may be easier to hold the top plate and turn the module over, lifting the housing off the top panel. If the Viton® gasket is loose, tack it down with 495 LocTite™ super bonder™. (Loctite Corp., USA) (or other suitable adhesive capable of adhering to the gasket material and to the aluminum case).
4www.PressureSystems.com