Pressure Systems, Inc. | 9016 Upgrade Instructions |
When performing any type of maintenance of NetScanner™ System components, the following guidelines and precautions should always be followed:
!Verify that the work area and technicians are properly grounded to prevent damage to electronic components due to electrostatic discharge.
!Ensure that all electrical and pneumatic connections have been removed from the module.
!Ensure that the work area is free of dust and other possible contaminants that may affect the high tolerance machined parts (and pneumatic seals, if model has an integral manifold).
!Care must be taken to prevent contaminants from reaching
!Never use sharp objects to cut tubing from the bulged tubes. The tiny scratches left on the tubes could cause leaks.
In the process of performing this upgrade, the following tools will be required:
!
!small cutters.
Module Disassembly
The following procedure should be used to disassemble your model 9016 prior to this upgrade.
(1)Place the scanner with its external connectors facing up. With one hand holding the module housing, remove all screws securing the top plate to the module housing. These are located around the perimeter of the top panel of the module housing. The 9016 uses twelve (12)
Use caution not to slip and scratch the top plate
3www.PressureSystems.com