23How to Use the VPAP Menus
SmartStart™
The VPAP has a function called SmartStart which can be enabled by your
clinician. If SmartStart is enabled, VPAP will start automatically when you
breathe into the mask and will stop automatically when you take your mask
off. This means you do not have to press the Front key to begin or end
treatment.
Note: If you select “Mir Full” as the mask option, SmartStop is
automatically disabled. SmartStart may not work with an Ultra Mirage Full
Face Mask due to safety features of the mask.
When Mask Alarm is set to ON, SmartStart/Stop automatically reverts to
OFF. SmartStop cannot be used with Mask Alarm because if a high leak
occurs, SmartStop will stop treatment before the Mask Alarm signal is
activated.
Settings for Mask Types
The following table shows the setting that should be selected for each mask
type.
Table 3: Settings for mask types
Humidifier NONE Selects the type of humidifier to
be used with the VPAP.
NONE, PASSOVER,
HUMIDAIRE
If the HumidAire 2i
is used, the above
setting options do
not appear and H2i
is displayed.
SmartStartOFF Turns the SmartStart function on
or off.
See “SmartStart™” on page 23.
ON/OFF
Mask Alarm ON
(VPAP III ST);
OFF
(VPAP III)
Turns the Mask Alarm signal on
or off.
If enabled, the Mask Alarm
feature will alert you when a
high mask leak is detected. An
audible tone will sound and a
high leak message will appear
on the LCD screen.
ON/OFF
Function Default Function Description Setting Options
Settings Mask
MIRAGE Mirage Nasal Mask
ULTRA Ultra Mirage Nasal Mask