Troubleshooting

Possible reason

Solution

Pillows won’t seal properly, are uncomfortable or cause red marks

English

Pillows may have been fitted incorrectly, incorrectly adjusted, or the headgear is too tight.

Inner wall position of one/ both pillows is incorrect.

Pillows are wrong size. Pillows may be dirty.

Carefully follow instructions in “Fitting your Swift LT”. Check headgear is not over-tightened.

Check ResMed logo on top of pillows is facing outwards.

Squeeze pillow sides to release inner wall. The correct position is shown in “Reassembling your Swift LT”.

Talk to your clinician.

Clean pillows according to instructions in “Cleaning your Swift LT in the Home”.

Nasal pillows system is too noisy

System is assembled

Reassemble the Swift LT. Make sure the pillows are

incorrectly.

fully inserted into the frame.

Before using your Swift LT

WARNINGS

The vent holes must be kept clear.

The mask should not be used unless the CPAP system is turned on and operating properly.

Follow all precautions when using supplemental oxygen.

Oxygen flow must be turned off when the flow generator is not operating, so that unused oxygen does not accumulate within the flow generator enclosure and create a risk of fire.

At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration varies, depending on the pressure settings, patient breathing pattern, mask, point of application and leak rate.

The technical specifications of the mask are provided for your clinician to check that they are compatible with the flow generator. If used outside specification or if used with incompatible devices, the seal and comfort of the mask may not be effective, optimum therapy may not be achieved, and leak, or variation in the rate of leak, may affect the flow generator function.

Discontinue using the Swift LT if you have ANY adverse reaction to the use of the mask, and consult your physician or sleep therapist.

As with all masks, some rebreathing may occur at low CPAP pressures.

Refer to your flow generator manual for details on settings and operational information.

Remove all packaging before using the mask.