Technical Specifications

Note: The mask system does not contain latex, PVC or DEHP materials.

Pressure–Flow Curve

 

70.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure

Flow

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(L/min)

60.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(cm H2O)

(± 6.0 L/min)

50.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4

22

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rate

40.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8

32

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow

30.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

12

41

20.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16

48

Vent

10.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

20

54

 

0.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

4

6

8

10

12

14

16

18

20

22

24

26

28

30

 

Mask Pressure (cm H2O)

Note: As a result of manufacturing variations, the vent flow rate may vary.

Dead Space Information

Dead space: the empty volume of the mask up to the swivel.

The dead space of the mask varies according to cushion size. It is 242 mL for the large size.

Standards

CE designation in accordance with EC directive 93/42/EEC, class IIa. ISO 14971, ISO 13485, ISO 10993-1 and ISO 17510-2.

Therapy Pressure

4 to 20 cm H2O

Resistance

Drop in pressure measured (nominal) at 50 L/min: 0.5 cm H2O

at 100 L/min: 1.6 cm H2O

Sound

The A-weighted sound power of the mask is 28 dBA.

Environmental Conditions

Operating temperature: +5°C to +40°C

Operating humidity: 10%–95% relative humidity non-condensing

Storage and transport: -20°C to +60°C

Storage and transport humidity: 10%–95% relative humidity non-condensing