ResMed C-Series IEC60601-1-2, Compliance, Electromagnetic, Immunity test, environment, test level

Models: C-Series

1 18
Download 18 pages 18.81 Kb
Page 14
Image 14
IEC60601-1-2

Guidance and manufacturer’s declaration—electromagnetic immunity

The C-Series Tango system is intended for use in the electromagnetic environment spec fied below. The customer or the user of the C-Series Tango system should assure that the device is used in such an environment.

 

IEC60601-1-2

Compliance

 

Electromagnetic

Immunity test

 

environment—

test level

level

 

 

 

guidance

 

 

 

 

 

 

 

 

 

Electrostatic

±6 kV contact

±6 kV contact

 

Floors should be wood,

discharge (ESD)

 

 

 

concrete or ceramic tile. If

IEC 61000-4-2

±8 kV air

±8 kV air

 

floors are covered w th

 

 

 

 

synthetic material, the

 

 

 

 

relative humid ty should be

 

 

 

 

at least 30%.

 

 

 

 

 

Electrical fast

±2 kV for power

±2 kV

 

Mains power qual ty should

transient/burst

supply lines

 

 

be that of a typical

IEC 61000-4-4

±1 kV for input/

Not Applicable

 

commercial or hosp tal

 

 

environment.

 

output lines

 

 

 

 

 

 

 

 

Surge

±1 kV d fferential

±1 kV d fferential

 

Mains power quality should

IEC 61000-4-5

mode

mode

 

be that of a typical

 

±2 kV common

±2 kV common

 

commercial or hosp tal

 

 

environment.

 

mode

mode

 

 

 

 

 

 

 

Voltage dips, short

<5% Ut

< 12V

 

Mains power quality should

interruptions and

(>95% d p in Ut)

(>95% d p in 240V)

 

be that of a typical

voltage variations on

for 0.5 cycle

for 0.5 cycle

 

commercial or hosp tal

power supply input

 

 

 

environment.

lines.

40% Ut

96V

 

If the user of the C-Series

IEC 61000-4-11

(60% dip in Ut)

(60% dip in 240V)

 

Tango system requires

 

for 5 cycles

for 5 cycles

 

continued operation during

 

70% Ut

168V

 

power mains interruptions,

 

 

it is recommended that the

 

(30% dip in Ut)

(30% dip in 240V)

 

C-Series Tango system be

 

for 25 cycles

for 25 cycles

 

powered from an

 

<5% Ut

<12V

 

uninterruptible power

 

 

source.

 

(>95%dip in Ut)

(>95%d p in 240V)

 

 

 

for 5 sec

for 5 sec

 

 

 

 

 

 

 

Power frequency

3 A/m

3 A/m

 

Power frequency magnetic

(50/60 Hz)

 

 

 

fields should be at levels

magnetic field

 

 

 

characteristic of a typical

IEC 61000-4-8

 

 

 

location in a typical

 

 

 

 

commercial or hosp tal

 

 

 

 

environment.

 

 

 

 

 

NOTE: Ut is the AC mains voltage prior to application of the test level.

 

 

 

 

 

 

12

Page 14
Image 14
ResMed C-Series manual IEC60601-1-2, Compliance, Electromagnetic, Immunity test, environment, test level, guidance