Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration varies, depending on the pressure settings, patient breathing pattern, mask, point of application and leak rate.

Do not use the S8 Auto 25 if there are obvious external defects or unexplained changes in performance.

Do not open the S8 case. There are no user serviceable parts inside. Repairs and internal servicing should only be performed by an authorised service agent.

Explosion hazard—do not use in the vicinity of flammable anesthetics.

Cautions

!At low pressures, the flow through the exhalation ports of your mask may not clear all exhaled gas from the tubing. Some rebreathing may occur.

The airflow for breathing produced by this device can be as much as 11ºF (6ºC) higher than the temperature of the room. Caution should be exercised if the room temperature is warmer than 90ºF (32ºC).

When AC mains power (100–240V AC) is not available, always use a ResMed DC/DC Converter 24V/50W. (The DC/DC Converter 24V/50W is available as an optional accessory. It is not supplied with all models.)

Note: The above are general warnings and cautions. Specific warnings, cautions and notes appear with the relevant instructions in the manual.

English

General Warnings and Cautions

17

Page 20
Image 20
ResMed S8 Auto 25 manual General Warnings and Cautions