Guidance and Manufacturer’s Declaration—Electromagnetic Emissions and Immunity

Guidance and manufacturer’s declaration—electromagnetic emissions

The S8 Auto 25 device is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment.

Emissions test

Compliance

Electromagnetic environment—guidance

 

 

 

RF emissions CISPR11

Group 1

The S8 device uses RF energy only for its internal

 

 

function. Therefore, its RF emissions are very low

 

 

and are not likely to cause any interference in

 

 

nearby electronic equipment.

 

 

 

RF emissions CISPR 11 with serial adapter

Class B

The S8 device is suitable for use in all

 

 

establishments, including domestic

RF emissions CISPR 11 with USB adapter

Class B

establishments and those directly connected to

 

 

the public low-voltage network that supplies

Harmonic Emissions IEC 61000-3-2

Class A

buildings used for domestic purposes.

 

 

 

Voltage Fluctuations/Flicker Emissions IEC

Complies

 

61000-3-3

 

 

 

 

 

Warnings: The S8 device should not be used adjacent to or stacked with other equipment.

If adjacent or stacked use is necessary, the S8 device should be observed to verify normal operation in the configuration in which it will be used. The use of accessories (eg, humidifiers) other than those specified in this manual is not recommended. They may result in increased emissions or decreased immunity of the S8 device.

Recommended separation distances between portable and mobile RF communications equipment and the S8 series of devices

The S8 device is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the S8 device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the S8 device as recommended below, according to the maximum output power of the communications equipment.

Separation distance according to frequency of transmitter (m)

Rated maximum output

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

power of transmitter (W)

d = 1.17 P

d = 0.35 P

d = 0.7 P

 

 

 

 

0.01

0.17

0.04

0.07

 

 

 

 

0.1

0.37

0.11

0.22

 

 

 

 

1

1.17

0.35

0.7

 

 

 

 

10

3.69

1.11

2.21

 

 

 

 

100

11.70

3.50

7.0

 

 

 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

18

Page 21
Image 21
ResMed S8 Auto 25 manual Separation distance according to frequency of transmitter m