Introduction

The S8TM Auto 25 self adjusting sleep apnoea system is indicated for the treatment of obstructive sleep apnoea (OSA) in patients > 30 kg. The S8 Auto 25 is intended for home and hospital use.

Contraindications

The S8 Auto 25 is not a life support device and may stop operating during power failure or certain fault conditions. It should not be used by patients who are dependent on continuous therapy.

Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions:

severe bullous lung disease

pneumothorax

pathologically low blood pressure

dehydration

cerebrospinal fluid leak, recent cranial surgery or trauma.

Adverse Effects

Patients should report unusual chest pain, severe headache or increased breathlessness to their prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment.

The following side effects may arise during the course of therapy with the device:

drying of the nose, mouth or throat

nosebleed

bloating

ear or sinus discomfort

eye irritation

skin rashes.

WARNING

!Read the entire manual before using this S8 device.

The S8 Auto 25 System

The S8 Auto 25 has two treatment modes, CPAP and VAuto, and comprises the following elements:

S8 Auto 25 device

2 m air tubing

Power cord

Travel bag

ResScan™ Data Card. Optional components include:

3 m air tubing

DC/DC Converter 24V/50W.

English

Introduction 1

Page 4
Image 4
ResMed manual Introduction, S8 Auto 25 System, Contraindications, Adverse Effects