Troubleshooting
Solution
Ensure the power cable is connected and the power outlet (if available) is on.
Fully insert the DC plug.

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open these devices.

Problem/Possible cause

No display

Power is not connected.

The DC plug is partially inserted into the back of the device or inserted too slowly.

Insufficient air delivered from the device

Ramp time is in use.

Wait for air pressure to build up or change ramp time.

Air filter is dirty.

Replace air filter.

Air tubing is not connected properly.

Check air tubing.

Air tubing is blocked, pinched or

Unblock or free the air tubing. Check the air tubing for

punctured.

punctures.

Mask and headgear are not positioned

Adjust position of mask and headgear.

correctly.

 

Incorrect air tubing selected.

If you are using the SlimLine, Standard or 3m air tubing

 

ensure that you have the correct air tubing selected via the

 

menu.

 

Device does not start when you breathe into the mask

 

 

Breath is not deep enough to trigger

Take a deep breath in and out through the mask.

SmartStart/Stop.

 

There is excessive leak.

Adjust position of mask and headgear.

Note: If Leak Alert is enabled, an

Air tubing not connected properly. Connect firmly at both

audible alert is activated and a high leak

ends.

message is displayed.

 

SmartStart/Stop is disabled.

Enable SmartStart/Stop.

Note: SmartStart/Stop is not available if

 

Leak Alert is enabled.

 