ResMed AutoSet CS, VPAP S, S9 Auto 25 manual Solution, Problem/Possible cause, English

Models: S9 Auto 25 VPAP S

1 15
Download 15 pages 59.63 Kb
Page 7
Image 7
Solution

Problem/Possible cause

Solution

 

Device does not stop when you remove your mask

 

 

SmartStart/Stop is disabled.

Enable SmartStart/Stop.

Note: SmartStart/Stop is not available if

 

Leak Alert is enabled.

 

SmartStart/Stop is enabled but the device does not stop automatically when you remove your mask

Incompatible mask system being used.

Only use equipment recommended by ResMed.

The patient is using a nasal pillows

Disable SmartStart/Stop.

mask or pediatric mask with a set

 

pressure less than 7 cm H2O.

 

Pressure rises inappropriately

 

 

 

Talking, coughing or breathing in an

Avoid talking with a nasal mask on, and breathe as normally

unusual manner.

as possible.

Mask cushion is buzzing against the

Adjust the headgear.

skin.

 

Cushion seated incorrectly causing

Adjust headgear or re-fit cushion.

excessive leak.

 

 

Displays error message: High temperature fault, refer to user manual

Device has been left in a hot

Allow to cool before re-use. Disconnect the power cord

environment.

and then reconnect it to restart the device.

Air filter is blocked.

Replace your air filter. Disconnect the power cord and then

 

reconnect it to restart the device.

Air tubing is blocked.

Check your air tubing and remove any blockages.

 

Disconnect the power cord and then reconnect it to restart

 

the device.

Humidifier setting is too high, resulting

Turn the humidifier setting down and empty the water from

in accumulation of water in the air

the air tubing.

tubing.

 

5 English

Page 7
Image 7