Alarm name | Cause | Solution | |
|
|
|
|
|
| The connected circuit is | Connect the correct |
|
| ||
|
| ||
|
| different from the one for | circuit. |
Change Circuit | which the device is |
| |
|
| configured. |
|
|
|
|
|
|
| The pressure line is | Reconnect the pressure |
|
| ||
|
| ||
|
| disconnected (see Figure | line. |
Prox. P. Lost | 9 on page 10). |
| |
|
|
|
|
|
| This alarm also indicates | Reconnect the patient |
|
| ||
|
| ||
|
| a wrongly connected or | circuit. |
Low Pressure | disconnected patient |
| |
|
| circuit. |
|
|
|
|
|
|
| A component in your | Clean, empty or replace |
|
| ||
|
| ||
|
| respiratory circuit is | the components of your |
High Pressure | blocked. | respiratory circuit. | |
|
|
| Contact your HME |
|
|
| provider if this alarm |
|
|
| persists. |
|
|
|
|
|
| Low inspired volume or | Check your circuit. |
|
| ||
|
| ||
|
| low expired volume. |
|
Low Vti or Low Vte |
|
| |
|
|
|
|
| High Vti | High inspired volume. | Check your circuit and |
| |||
| |||
|
|
| make sure there are no |
|
|
| leaks. |
|
|
| |
Table 4: Technical alarms (not an exhaustive list) |
| ||
|
|
|
|
Alarm name | Cause | Solution | |
|
|
|
|
| Tech [n] | Technical alarm. The | Contact your HME |
| |||
| |||
|
| alarm number indicates | provider and give the |
|
| the type of incident. | number of the alarm. |
|
|
|
|
| Turbine | Alarm indicating that the | Contact your HME |
| |||
| |||
|
| turbine has stopped. | provider. |
|
|
|
|
|
| The temperature of the | Contact your HME |
|
| ||
|
| ||
|
| internal battery is too | provider. |
Temp Out Of | high or too low. |
| |
Range |
|
| |
|
|
|
|
|
| The charge level of the | Connect the device to |
|
| ||
|
| ||
|
| backup battery (internal | the mains power supply |
Check Date&Time | clock memory) is low. | without delay and check | |
|
|
| the date and time. |
|
|
|
|