ALARMS & SAFETY FUNCTIONS
This unit has the alarms and safety functions shown in Table 5, and also self diagnostic functions.
Table 5 Alarms and safety functions
Alarm & Safety | Situation | Indication | Buzzer | Safety operation | |
High temperature | If the chamber temperature is higher | ALARM lamp is flashed. | Intermittent tone with | Remote alarm with 15 | |
than the temperature at which the | Temperature indicator is | ||||
alarm | 15 minutes delay. | minutes delay. | |||
high temperature alarm is activated. | flashed. | ||||
|
|
| |||
Low temperature | If the chamber temperature is lower | ALARM lamp is flashed. | Intermittent tone with | Remote alarm with 15 | |
than the temperature at which the | Temperature indicator is | ||||
alarm | 15 minutes delay. | minutes delay. | |||
low temperature alarm is activated. | flashed. | ||||
|
|
| |||
| In the case of power failure. |
|
|
| |
Power failure alarm | When power switch is turned OFF. | ALARM lamp is flashed. | Intermittent tone | Remote alarm. | |
When the power to the unit is | |||||
|
|
|
| ||
| disconnected. |
|
|
| |
When there is no key pressing in | Chamber temperature is | Finishing of each | |||
each setting mode for 90 seconds. | displayed. |
| setting mode. | ||
|
| ||||
Key lock | When the key lock is “ON”. | Change of setting is | |||
disable. | |||||
|
|
|
| ||
Thermal sensor | If the thermal sensor is | ALARM lamp is flashed. |
| Remote alarm. | |
E01 and chamber temp. are | Intermittent tone | ||||
Abnormality | disconnected. | Continuous running. | |||
displayed alternately. |
| ||||
(for |
|
|
| ||
| ALARM lamp is flashed. |
|
| ||
upper chamber | If the thermal sensor is |
| Remote alarm. | ||
E02 and chamber temp. are | Intermittent tone | ||||
only) | Continuous running. | ||||
displayed alternately. |
| ||||
|
|
|
| ||
Thermal sensor | If the thermal sensor is | ALARM lamp is flashed. |
| Remote alarm. | |
E03 and chamber temp. are | Intermittent tone | ||||
abnormality | disconnected. | Continuous running. | |||
displayed alternately. |
| ||||
|
|
| |||
| ALARM lamp is flashed. |
|
| ||
lower chamber | If the thermal sensor is |
| Remote alarm. | ||
E04 and chamber temp. are | Intermittent tone | ||||
only) | Continuous running. | ||||
displayed alternately. |
| ||||
|
|
|
| ||
| If the protective sensor for | ALARM lamp is flashed. |
|
| |
Thermal sensor | E05 and chamber temp. are |
|
| ||
compressor is disconnected. |
| Remote alarm. | |||
abnormality | displayed alternately. | Intermittent tone | |||
| |||||
If the protective sensor for | ALARM lamp is flashed. | Normal operation. | |||
| |||||
only) | compressor is | E06 and chamber temp. are |
|
| |
| displayed alternately. |
|
| ||
|
|
|
| ||
| If the thermal sensor for lower | ALARM lamp is flashed. |
|
| |
| E07 and chamber temp. are |
|
| ||
Thermal sensor | chamber is disconnected. |
| Remote alarm. | ||
displayed alternately. | Intermittent tone | ||||
abnormality |
| ||||
| ALARM lamp is flashed. | Normal operation. | |||
If the thermal sensor for lower |
| ||||
E08 and chamber temp. are |
|
| |||
| chamber is |
|
| ||
| displayed alternately. |
|
| ||
|
|
|
| ||
Battery switch | When battery switch is OFF at the | ALARM lamp is flashed. | Intermittent tone | Remote alarm. | |
check | time of alarm test. | E09 is flashed. | |||
|
| ||||
| In the case of failure of compressor |
|
| Remote alarm. | |
|
|
| Compressor stops | ||
| cooling fan motor. |
|
| ||
|
|
| running when the temp | ||
Compressor temp. | In the case of abnormal high |
|
| ||
E10 and chamber temp. are | Intermittent tone | reaches about 100oC. | |||
Abnormality (MDF- | temperature due to the dust on the | displayed alternately. | Electromagnetic valve | ||
U537, U537D only) | condenser. |
| |||
|
| 1, 2 and compressor | |||
| In the case of abnormal high |
|
| ||
|
|
| stop. | ||
| ambient temperature. |
|
| ||
|
|
| |||
|
|
|
|
Note:
1.The above power failure alarm is available when the battery switch is ON and the battery is charged. If the battery switch is OFF or the battery is discharged, only the remote alarm is activated.
2.The power failure alarm can be kept about 12 hours with the battery charged completely.
3.The chamber temperature is displayed for 5 seconds if the BUZZER key is depressed during the power failure alarm. At the same time, the alarm stops.
4.The remote alarm is silenced by pressing BUZZER key as the remote alarm is operated in conjunction with the buzzer, except for the power failure alarm.
5.After power failure, the operation is resumed with the condition before power failure since the temperature setting and alarm temperature setting are memorized in a nonvolatile memory.