Sanyo MDF-U537 Temperature Setting, Chamber temperature, Description of operation, Key operated

Models: MDF-U333 MDF-U537

1 30
Download 30 pages 54.07 Kb
Page 20
Image 20
TEMPERATURE SETTING

TEMPERATURE SETTING

Chamber temperature

Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the sequence indicated in the table. The example in the table is based on the assumption that the desired temperature is -25oC.

Note: The unit is set at the factory that the chamber temperature -30oC.

Table 1. Basic operation sequence (Example: Chamber temperature -25oC)

 

Description of operation

Key operated

Indication after operation

1

Connect the power plug to

----

The current chamber temperature is

exclusive power supply socket.

displayed.

 

 

 

(MDF-U537D only)

TOP

Top compartment lamp lights and the

2

By pressing the compartment select

current chamber temperature of top

BOTTOM

 

key, select the top compartment.

compartment is displayed.

 

 

3

Press SET key.

SET

The second digit of the temperature

indicator flashes.

 

 

 

 

Set the temperature to 25 with the

 

By pressing the key, settable digits

 

 

moves.

4

digit shift key and the numeric

 

 

By pressing the key, the numerical value

 

value shift key.

 

 

 

of the settable digits increases.

 

 

 

5

Press SET key.

SET

When pressed, the figure of settable

digit increases.

 

By pressing the compartment select

 

 

TOP

Set temperature is memorized and the

6

key, select the bottom

BOTTOM

current chamber temperature is displayed.

 

compartment. (MDF-U537D only)

 

 

 

 

(MDF-U537D only)

 

 

7

Perform the same operation as in the case of steps

The chamber temperature setting is stored, and the

3 to 5, and set the temperature for the bottom

current chamber temperature is displayed.

 

 

compartment.

 

 

Note:

Although the value of the chamber temperature setting can range from -18oC to -40oC, the guaranteed temperature when there is no load is -30oC when the external temperature is 30oC.

Note that defrosting of this unit is started manually (by pressing the DEF key for about 5 seconds) and stopped manually.

19

Page 20
Image 20
Sanyo MDF-U537, MDF-U333 Temperature Setting, Basic operation sequence Example Chamber temperature -25oC, Key operated