Alarm temperature setting

TEMPERATURE SETTING

Alarm temperature setting

This unit is provided with both high and low temperature alarms. The temperature at which the alarm is activated may be changed.

The available set range for high temperature alarm is between +5oC and +15oC, and -5oC and -15oC for low temperature alarm against the chamber temperature.

Note: The temperature alarm is set at ±10oC of the set temperature at the factory.

Display

Mode

Function

 

High temperature alarm set

See Table 5 on page 22

 

Low temperature alarm set

See Table 5 on page 22

 

High temperature alarm set for lower chamber of MDF-U537D

See Table 5 on page 22

 

Low temperature alarm set for lower chamber of MDF-U537D

See Table 5 on page 22

As an example, Table 3 shows the procedure to set the high temperature alarm so that the alarm can activate when the chamber temperature is 5oC higher than the set temperature.

Table 4 shows the procedure to set the low temperature alarm so that the alarm can activate when the chamber temperature is 5oC lower than the set temperature.

Table 3. Procedure for setting high temperature alarm

Description of operation

Key operated

1

----

 

2 Press Table 3. Procedure for setting high temperature alarm key for 5 seconds.

3

Press key and scroll the figure

 

to 1.

 

 

 

4

Press SET key.

SET

5

Set the temperature to 005 with the

 

key and key.

 

 

 

6

Press SET key.

SET

Indication after operation The current chamber temperature is displayed.

The first digit is flashed.

When pressed, the figure of settable digit increases.

The right digit is flashed.

Pressing the key shifts the digit which can be set. When pressed, the figure of settable Table 4. Procedure for setting low temperature alarm digit increases.Manual background Alarm temperature is memorized and

the current chamber temperature is displayed.

Table 4. Procedure for setting low temperature alarm

Description of operation

Key operated

1

----

 

2 Press Manual background key for 5 seconds.

3

Press key and scroll the figure

 

to 2.

 

 

 

4

Press SET key.

SET

5

Set the temperature to -05 with the

 

key and key.

 

 

 

6

Press SET key.

SET

Indication after operation The current chamber temperature is displayed.

The right digit is flashed.

When pressed, the figure of settable digit increases.

The right digit is flashed.

Pressing the key shifts the digit which can be set. When pressed, the figure of settable Manual background digit increases.Manual background Alarm temperature is memorized and

the current chamber temperature is displayed.

21

Page 22
Image 22
Sanyo MDF-U537, MDF-U333 Alarm temperature setting, Temperature Setting, Procedure for setting high temperature alarm