OPERATING INSTRUCTIONS

Stable running

Table below shows the basic procedure for setting the chamber temperature. Perform key operations in the sequence indicated in the table. The example in the table is based on the assumption that the desired temperature is 50°C and continuous running.

50°C

Time

Basic operation sequence (Example: Chamber temperature; 50 C, Continuous running)

 

Description of operation

Key operated

 

Indication after operation

 

 

 

 

 

 

О ALARM

О DELAY

 

 

 

 

 

 

1

Press the power key.

 

POWER

 

О SET VALUE

О START

 

 

 

 

 

 

0 MODE 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(The current chamber temperature is displayed.)

 

 

 

 

 

 

 

 

 

 

 

 

0 ALARM

• DELAY

 

 

 

 

 

 

2

(Delay time reset)

 

DELAY

 

• SET VALUE

О START

0

0

 

0

0

 

Press the delay timer key.

 

 

О MODE 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(The fourth digit of the indicator flashes.)

 

 

 

 

 

 

 

 

Press the delay timer key to reset

 

 

 

 

О ALARM

• DELAY

 

 

 

 

 

 

3

 

DELAY

 

О MODE 1

О START

 

 

 

0

0

 

 

 

 

 

 

 

• SET VALUE

 

 

 

 

 

 

 

the delay time.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press the delay timer key and the

 

 

 

 

О ALARM

О DELAY

 

 

 

 

 

 

4

 

DELAY

О SET VALUE

О START

 

 

 

E5

 

current chamber temp, is displayed.

 

О MODE 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

О ALARM

О DELAY

 

 

 

 

 

 

к

(Temperature set)

 

CALL

 

• SET VALUE

О START

 

0

 

0

0

 

Press the call key.

 

 

 

 

 

 

 

 

 

 

 

(The third digit of the indicator 1lashes.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set the temp, to 050 with the digit

 

 

 

 

О ALARM

О DELAY

 

 

 

 

 

 

6

shift key and the numeric value

 

 

 

 

• SET VALUE

О START

 

0

5

0

 

 

 

 

 

0 MODE 1

 

 

 

 

 

shift key.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

О ALARM

О DELAY

 

 

 

 

 

 

7

(Time set)

 

CALL

 

• SET VALUE

О START

a

о

 

0

U

 

Press the call key.

 

 

0 MODE 1

 

 

 

 

 

 

(The fourth digit of the indicator flashes.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set the time to --:- with the digit

_

 

 

О ALARM

О DELAY

-

I - -

 

 

 

 

• SET VALUE

О START

8

shift key and numeric value shift

 

 

 

 

 

 

 

 

 

 

 

 

0 MODE 1

 

 

 

 

 

 

 

key.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(Start running)

 

 

 

 

0 ALARM

О DELAY

 

 

 

 

 

 

 

 

RUN

 

0 SET VALUE

• START

 

 

 

E

с

 

 

 

 

 

 

 

 

 

 

 

 

9

 

 

 

 

 

 

 

 

 

 

 

Press the run/stop key.

 

STOP

 

О MODE 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(The unit runs.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Note:

 

 

 

 

 

 

 

 

 

 

 

 

 

When the delay time is not set, operation 2, 3, and 4 is not necessary.

-:— of the indicator is displayed by the numerical value shift key at fourth digit flashing.

Page 15
Image 15
Sanyo MOV-112S, MOV-212S instruction manual Stable running