ALARMS & SAFETY FUNCTIONS
This unit has the alarms and safety functions shown in table below, and also self diagnostic functions.
Alarms and safety functions
Alarm & | Situation | Indication | Buzzer | Safety operation | |
Safety | |||||
|
|
|
| ||
| If the internal temperature |
|
|
| |
Automatic set | deviates from the set | Alarm lamp lights. |
| — | |
temperature | temperature by +10°C or more | All digits on the | Intermittent tone | ||
Alarm | (and if timer mode 1, by | indicator flash. |
|
| |
| more) |
|
|
| |
| If a key operation is not |
| — | The setting mode | |
Auto return | performed for about 45 sec. in | Normal display mode. | |||
is cancelled. | |||||
| each setting mode. |
|
| ||
|
|
|
| ||
Key lock switch | When the key lock switch is |
| — | Key input is | |
turned ON. |
| disabled. | |||
|
| ||||
Temperature | If the temperature sensor goes | Alarm lamp lights. |
|
| |
sensor | E01 is displayed on | Intermittent tone | Heater OFF | ||
open circuit. | |||||
abnormality | the indicator. |
|
| ||
|
|
| |||
|
| Alarm lamp lights. |
|
| |
| If the triac goes open circuit. | E02 is displayed on | Intermittent tone | Heater OFF | |
Triac |
| the indicator. |
|
| |
abnormality |
| Alarm lamp lights. |
|
| |
| If the triac goes short circuit. | E03 is displayed on | Intermittent tone | Heater OFF | |
|
| the indicator. |
|
| |
|
| Alarm lamp lights. |
|
| |
| If the relay goes short circuit. | E04 is displayed on | Intermittent tone | Heater OFF | |
Relay |
| the indicator. |
|
| |
If the relay goes open circuit or |
|
|
| ||
abnormality | Alarm lamp lights. |
|
| ||
fan motor and heater goes short |
|
| |||
| E05 is displayed on | Intermittent tone | Heater OFF | ||
| circuit. Or the ambient temp, | the indicator. |
|
| |
| of circuit is 65°C or more. |
|
| ||
|
|
|
| ||
| (when unit is not running) |
|
|
| |
| The security circuit is activated | Without display |
| Heater OFF | |
| by independent temp, sensor if | Continuous tone | forcedly by | ||
| change. | ||||
| the chamber is abnormal over- |
| external circuit. | ||
Independent |
|
| |||
heating. |
|
|
| ||
|
|
| |||
(When unit is running) |
|
|
| ||
protection |
| Continuous tone |
| ||
The security circuit is activated | Alarm lamp lights. | Heater OFF | |||
| (When the temp, | ||||
| by independent temp, sensor if | E05 is displayed on | forcedly by | ||
| s decreased, | ||||
| the chamber is abnormal over- | the indicator. | external circuit. | ||
| ntermittent tone) | ||||
| heating. |
|
| ||
|
|
|
|
*The buzzer tone resulting from the independent