
18 
Tag Profiler
Read and Write power settings  can be optimized using the Tag Profiler . Before running the 
Tag Profiler, it is important to ensure that the  proper Gap Sensing procedure has been 
followed. It is recommended that you calibra te the printer to ensure that gap sensing settings 
are correct. 
To use the Tag Profiler, first ensure that the t ag type that you have selected from the Tag Type 
menu  or page 3 [GL4xxe] for a list of suppor ted tag types) either matches the tag type you a re 
about to calibrate or is at least of the sam e Class and data length (i.e. 64 or 96 bits). Next, check 
that Custom Tag is set to Duplicate. The correct defaults w ill then be set for Custom Tag Class 
and Tag (EPC Byte) Length. 
Before initiating the calibration cycle, the Tag Profiler can be optimized by setting appropriate 
limits on the following four custom  entries: 
•  Custom (Scan) Start Pos. Identifies the starting position of the calibrat ion scan. The default 
will start at the Top Of Form (TOF). To avoid in efficient scanning at points far away from the 
target tag, set the (Scan) Start Pos within one  inch of the center of the physical tag. E.g., if 
the center of the tag is physically 76mm f rom the TOF, set (Scan) Start Pos to 51mm (25mm 
(1 inch) before the center o f the tag). 
•  Custom Scan Length. This refers to the distance that the Tag Profiler will sc an to determine 
the optimum tag position. For improved pe rformance, set Custom Scan Length to 51mm or 
less. This will prevent the printer from look ing for a tag far beyond its actual location. 
•  Custom (Cal) Min Power. Sets the l ower level that will be tested during  calibration. To 
speed up calibration, set Custom Min Po wer to two points lower than the Custom Read 
Power that was set prior to initiating calibration. 
Introduction to the RFID configuration options