SECTION 2—COMPONENT REPLACEMENT

SECTION 2—COMPONENT

REPLACEMENT

CAUTION

Ensure not to damage ribbon cable when removing or installing front cover. DO NOT use Methyl Ethyl Ketone (MEK). This will damage the plastic.

Replacing the Gross Particle Filter or Filter Cover

NOTE: For this procedure, refer to

FIGURE 2.1.

NOTE: The gross particle filter should be replaced as needed.

1.Lift filter cover slightly and pull down to remove tabs from the grooves.

2.Lift out the filter from the filter cover.

3.Use a vacuum cleaner or wash with a mild liquid dish detergent (such as Dawn) and water. Rinse thoroughly.

4.Thoroughly dry the filter and inspect for fraying, crumbling, tears and holes.

5.Replace filter cover and/or filter.

6.Reinstall the filter cover by placing the tabs in the slots and engaging the magnet.

NOTE: Use only Invacare gross particle filter.

L

MagnetCover

 

 

Tabs

Filter

 

 

 

Grooves

Cover

 

Cover

Magnet

Filter

FIGURE 2.1 Replacing the Gross Particle

Filter or Filter Cover

Accessing XPO2

WARNING

When accessing the XPO2 remember that battery power is still being applied to the internal (PCB) printed circuit board. Care MUST be taken to minimize contact with the PCB. DO NOT use any kind of tool (screwdriver, pliers, etc.) when working on the internal structure, otherwise, injury and/or damage may occur. Battery power remains on the PCB as long as the battery connection is made to the PCB.

NOTE: For this procedure, refer to FIGURE 2.2 and FIGURE 2.3 on page 10.

Part No 1150748

9

XPO2

Page 9
Image 9
Shop-Vac XPO100B service manual Component Replacement, Accessing XPO2