A | Return the ‘REL’ Status |
|
|
A0 | Turn the ‘REL’ off | N | Return the ENBW setting |
A1 | Turn the ‘REL’ on | N0 | Select 1 Hz ENBW |
|
| N1 | Select 10 Hz ENBW |
B | Return Bandpass Filter Status |
|
|
B0 | Take out the Bandpass Filter | O | Return Offset Status |
B1 | Put in the Bandpass Filter | O0 | Turn off Offset |
|
| O1,v | Turn on Offset, v = offset |
C | Return the Reference LCD Status |
|
|
C0 | Display the Reference Frequency | P | Return the Phase Setting |
C1 | Display the Reference Phase Shift | Pv | Set the Phase to v. Abs(v) < 999 deg |
D | Return Dynamic Reserve Setting | Q | Return the value shown on the Output |
D0 | Set DR to LOW range |
| LCD |
D1 | Set DR to NORM range |
|
|
D2 | Set DR to HIGH range | R | Return the trigger mode |
|
| R0 | Set the trigger for rising edge |
En | Return Expand Status | R1 | Set the trigger for + zero crossing |
En,0 | Turn Expand off | R2 | Set the trigger for falling edge |
En,1 | Turn Expand on |
|
|
|
| S | Return the display status |
F | Return the Reference Frequency | S0 | Display X = AcosØ |
|
| S1 | Display Offset setting |
G | Return the Sensitivity Setting | S2 | Display Noise |
G1 | Select 10 nV |
|
|
... | T1 | Return | |
G24 | Select 500 mV | T1,1 | Set the |
|
| ... |
|
H | Return Preamp Status (1=installed) | T1,11 | Set the |
I | Return the Remote/Local Status | T2 | Return the |
I0 | Select Local: Front panel active | T2,0 | Remove post filter |
I1 | Select Remote: Front panel inactive | T2,1 | Set the post filter TC to 0.1 S |
I2 | Select Remote with full | T2,2 | Set the post filter TC to 1.0 S |
J | Set RS232 | V | Return the value of the SRQ mask |
Jn,m,o,p | Set | Vn | Set the SRQ Mask to the value n |
|
|
| (See the Status Byte definition) |
K1 | Simulates |
|
|
... | (see | W | Return the RS232 wait interval |
K32 | Simulates | Wn | Set RS232 wait interval to nX4mS |
L1 | Return Status of Line Notch Filter | Xn | Return the voltage at the rear panel |
L1,0 | Remove Line Notch Filter |
| analog port n. (n from 1 to 6) |
L1,1 | Insert Line Notch Filter |
|
|
|
| X5,v | Set analog port 5 to voltage v |
L2 | Return Status of 2XLine Filter | X6,v | Set analog port 6 to voltage v |
L2,0 | Remove 2XLine Notch Filter |
|
|
L2,1 | Insert 2XLine Notch Filter | Y | Return the Status Byte value |
|
| Yn | Test bit n of the Status Byte |
M | Return the f/2f Status |
|
|
M0 | Set reference mode to f | Z | Reset to default settings and cancel |
M1 | Set reference mode to 2f |
| all pending command |
6