Vpsd = cos(wr+Ø) cos(wst)
=1/2 cos[(wr + ws)t+Ø] + 1/2 cos[(wr - ws)t+Ø]
The sum frequency component is attenuated by the low pass filter, and only those difference frequency components within the low pass filter's narrow bandwidth will pass through to the dc amplifier. Since the low pass filter can have time constants up to 100 seconds, the
For signals which are in phase with the reference, the phase control is usually adjusted for zero phase difference between the signal and the reference. This can be done by maximizing the output signal. A more sensitive technique would be to adjust the phase to null the signal. This places the reference oscillator at 90 degrees with respect to the signal. The phase control can now be shifted by 90 degrees to maximize the signal. Alternatively, since the phase control is well calibrated, the phase of the signal can be measured by adding 90 degrees to the phase setting which nulls the signal.
Understanding the Specifications
The table below lists some specifications for the SR510
We have chosen a reference frequency of 5 kHz so as to be in a relatively quiet part of the noise spectrum. This frequency is high enough to avoid low frequency '1/f' noise as well as line noise. The frequency is low enough to avoid phase shifts and amplitude errors due to the RC time constant of
the source impedance and the cable capacitance.
The
A
single pole filter is 1/4RC. The output will converge exponentially to the final value with a 10 second time constant. If we wait 50 seconds, the output will have come to within 0.7% of its final value.
The dynamic reserve of 60 dB is required by our expectation that the noise will be a thousand times larger than the signal. Additional dynamic reserve is available by using the bandpass and notch filters.
A
Specifications for the Example Measurement
Specification | Value | Error |
Full Scale Sensitivity | 100 nV |
|
Dynamic Reserve | 60 dB |
|
Reference Frequency | 5 kHz |
|
Gain Accuracy | 1% | 1% |
Output Stability | 0.1%/°C | 1% |
< 7 nV/√ Hz 1.2% | ||
Output Time Constant | > 10 S | 0.7% |
Total RMS Error |
| 2% |
Shielding and Ground Loops
In order to achieve the 2% accuracy given in this measurement example, we will have to be careful to minimize the various noise sources which can be found in the laboratory. (See Appendix A for a brief discussion on noise sources and shielding) While intrinsic noise (Johnson noise, 1/f noise and alike) is not a problem in this measurement, other noise sources could be a problem. These noise sources can be reduced by proper shielding.
There are two methods for connecting the
In the first method, the
a
25