C l e a n i n g

9

11. If an aerosol mask is used, place the mask over

 

mouth and nose (Fig. 8). As aerosol begins to flow,

 

inhale deeply and slowly through mouth, then

 

exhale slowly.

 

12. When treatment is complete, turn unit off by press-

 

ing the power switch to the “Off” position. Unplug

 

the unit from the wall power outlet.

 

NEBULIZER CLEANING

 

All parts of the nebulizer, except tubing, should be cleaned

Figure 7

according to the following instructions. Your physician

 

and/or Sunrise provider may specify a certain cleaning proce-

 

dure. If so, follow their recommendations.

 

To prevent possible risk of infection from contami- nated medication, cleaning of the nebulizer is rec- ommended after each aerosol treatment.

Disinfecting is recommended once a day.

Clean After Every Use:

1.With power switch in the “Off” position, unplug power cord from wall outlet.

2.Disconnect tubing from the air-inlet connector and set aside.

3.Disassemble mouthpiece or mask from cap. Open nebulizer by turning cap counterclockwise and removing baffle.

4.Wash all items, except tubing, in a hot water/dish- washing detergent solution. Rinse under hot tap water for 30 seconds to remove detergent residue. Allow to air dry.

5.For reusable nebulizer only, clean daily in dishwash- er using top shelf.

CAUTION– The DeVilbiss Reusable Nebulizer is dishwasher safe, but do not place any DeVilbiss Disposable Nebulizer parts in an automatic dishwasher; doing so could cause damage.

Figure 8

Figure 9

A-655

Page 9
Image 9