12
S p e c i f i c a t i o n s
SPECIFICATIONS
Size 3655 Series | 7.5" W x 4.0" H x 7.2" D (19.1 cm x 10.2 cm x 18.3 cm) | |
4650 Series | 7" W x 3.8" H x 13" D (18 cm x 10 cm x 33 cm) | |
5650 Series | 10.1" W x 10.5" H x 6.5" D (26 cm x 27 cm x 17 cm) | |
Weight 3655 Series | 4.8 lbs. (2.18 kg) | |
4650 Series | 5.3 lbs. (2.40 kg) | |
5650 Series | 7.1 lbs. (3.2 kg) | |
Electrical |
|
|
Requirements 3655D, 3655T, 4650D & 46502 | 2.5 amps, 115 VAC (±10%) 60 Hz | |
3655I & 4650I | 1.4 amps, 230 VAC (±10%) 50 Hz | |
3655K & 3655P | 1.4 amps, 230 VAC (±10%) 60 Hz | |
5650D, 5650H & 5650C | 1.3 amps, 115VAC 60Hz | |
5650I | 0.8 amps, 230VAC 50Hz | |
5650K & 5650P | 0.8 amps, 230VAC 50/60Hz | |
Sound Level 3655 Series | 58 dBA | |
4650 Series | 53 dBA | |
5650 Series | 51 dBA | |
Power Consumption 3655D, 3655T, 4650D & 46502 | 140 watts maximum | |
3655I, 3655K, 3655P & 4650I | 160 watts maximum | |
5650D, 5650H & 5650C | 90 watts maximum | |
5650I, 5650K & 5650P | 100 watts maximum | |
Maximum Compressor Pressure 3655 & 4650 series | 35 psig or greater (241 kPa)* | |
| 5650 series | 30 psig or greater (206 kPa)* |
Compressor Free Air Flow 3655 & 4650 | 8 lpm or greater* | |
Operating | 5650 | 9 lpm or greater* |
|
| |
Temperature Range all Models except 5650C | +40° to +104°F (+5 to +40°C) | |
5650C | +40° to +101°F (+5 to +38.5°C) | |
Operating Humidity all Models | up to 95% | |
Storage/Transport Temperature Range all Models | ||
Storage/Transport Humidity all Models | up to 95% | |
Supplied Nebulizer Capacity all models | 6 ml (cc) | |
Supplied Nebulizer Nebulization Rate all models | 0.15 ml/min or greater* | |
Supplied Nebulizer MMAD all models | 5 microns or less** |
*Conditions may vary based on altitude above sea level, barometric pressure, and temperature. **Value determined with minimum flow rate through the nebulizer (20°C, sea level).
There are no hazardous materials used in the construction of this unit, therefore, there are no known risks associated with disposal of the equipment.
This product (models 3655D & 3655T) complies with the following electromagnetic compatibility standard: IEC
3655D & 3655T Equipment Classifications: |
|
Protection against electric shock: | Class II |
Degree of protection against electric shock: | Type BF |
Equipment not suitable for use in the presence of flammable anaesthetic mixture with
air or with oxygen or nitrous oxide |
|
Mode of operation: | Continuous operation |