HP 16532A - Performance Tests

3-12. TIME MEASUREMENT ACCURACY

This test uses a precise frequency source to check the accuracy of time measurement functions.

Specification:

±[(0.005% x Δt) + (2 x 106 x delay setting) + 150 ps]

Equipment Required:

The following equipment is required for this test. Procedures are based on the model or part number rec- ommended.

 

 

Recommended

Equipment Required

Critical Specification

Model/Part Number

 

 

 

Signal Generator

200 MHz, timebase accuracy 0.25 ppm

HP 8656B Opt. 001

Cable

50 Ω BNC (m to m) 48-inch

HP 10503A

Adapter

50 Ω Type N (m) to BNC (f)

HP 1250-0780

 

 

 

 

 

 

Procedure:

1.Set the signal generator to 181.81818 MHz (5.5 ns period) and approximately 600 mV rms.

2.Use the N-to-BNC adapter to connect the cable to the signal generator. Connect the free end of the BNC cable to the channel 1 input.

3.In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Set the following parameters in the order given:

 

Menu

 

Selection

 

Setting

 

 

 

 

 

 

Channel

 

Input

 

1

 

 

 

Probe

 

1:1

 

 

v/Div

 

400 mV

 

 

Offset

 

0.0 V

 

 

Coupling

 

50

Ω / DC

 

 

s/Div

 

2.00 ns

Trigger

 

Mode

 

Edge

 

 

Source

 

Chan 1

 

 

Level

 

0.0 V

 

 

Slope

 

Positive

Display

 

Mode

 

Normal

 

 

Grid

 

On

 

 

 

 

 

 

 

 

 

 

 

 

 