HP 16532A - Calibration
4-10. Equipment Required for Calibration
The equipment listed below is required for calibration: |
| |
Cable (2) | 50Ω BNC, | HP 10502A |
Cable | 50Ω BNC - SMB | HP |
Adapter | 50Ω BNC tee (m)(f)(f) | HP |
Adapter | 50Ω BNC (m)(m) | HP |
4-11. Loading the Default Calibration Factors
Once the default calibration factors are loaded, all calibrations must be done. This includes all of the calibrations in the Self Cal menu. The calibration must be performed in the exact sequence listed in table
Note
The calibration PROTECT/UNPROTECT switch on the oscilloscope board must be set to UNPROTECT.
Load the default calibration factors as follows:
1.Touch the menu field (top row, second from the left), then select Calibration from the
2.Touch the Mode field, then select Service Cal from the
3.Touch the Procedure field, then select Default Values from the
Note
After you touch the Start field, you can abort the calibration procedure either by touching the Cancel field, if it is available in the menu, or by touching either the Mode or Procedure fields if the continue field is still displayed on the screen.
4.Touch the Start field and follow the instructions on the display. The display should now look similar to the display shown in figure