HP 1GSa/s Digitizing Oscilloscope 16532A manual Equipment Required for Calibration

Models: 1GSa/s Digitizing Oscilloscope 16532A

1 90
Download 90 pages 21.04 Kb
Page 52
Image 52

HP 16532A - Calibration

4-10. Equipment Required for Calibration

The equipment listed below is required for calibration:

 

Cable (2)

50Ω BNC, 9-inch (equal length)

HP 10502A

Cable

50Ω BNC - SMB

HP 16532-61601

Adapter

50Ω BNC tee (m)(f)(f)

HP 1250-0781

Adapter

50Ω BNC (m)(m)

HP 1250-1236

4-11. Loading the Default Calibration Factors

Once the default calibration factors are loaded, all calibrations must be done. This includes all of the calibrations in the Self Cal menu. The calibration must be performed in the exact sequence listed in table 4-1.

Note

The calibration PROTECT/UNPROTECT switch on the oscilloscope board must be set to UNPROTECT.

Load the default calibration factors as follows:

1.Touch the menu field (top row, second from the left), then select Calibration from the pop-up menu.

2.Touch the Mode field, then select Service Cal from the pop-up menu.

3.Touch the Procedure field, then select Default Values from the pop-up menu.

Note

After you touch the Start field, you can abort the calibration procedure either by touching the Cancel field, if it is available in the menu, or by touching either the Mode or Procedure fields if the continue field is still displayed on the screen.

4.Touch the Start field and follow the instructions on the display. The display should now look similar to the display shown in figure 4-3.

Figure 4-3. Default Calibration Display

4-4

Page 52
Image 52
HP 1GSa/s Digitizing Oscilloscope 16532A manual Equipment Required for Calibration, Loading the Default Calibration Factors