HP 16532A - Performance Tests

3-7. AC CAL OUTPUT Test

The AC CAL OUTPUT port on the rear panel is used for instrument calibration and probe compensation. Though calibrator accuracy is not specified in the performance specifications, it must be within limits in or- der to provide accurate calibration.

Test Limits:

1000 Hz ± 10%, 0.8 Vp_p ± 10%

Equipment Required:

The following equipment is required for the test.

 

 

Recommended

Equipment Required

Critical Specification

Model/Part Number

 

 

 

Cable

50 Ω BNC (m to m) 48-inch

HP 10503A

 

 

 

 

 

 

Procedure:

1.Use the BNC cable to connect the AC CAL OUTPUT to channel 1 input of the oscilloscope module.

2.On the oscilloscope module set the following parameters in the order given:

 

Menu

 

Selection

Setting

 

 

 

 

 

Calibration

 

Mode

 

Service Calibration

 

 

Procedure

 

Oscillator Output

 

 

Signal

 

Probe Compensation

Channel

 

Coupling

 

1 MΩ/DC

 

 

Probe

 

1:1

 

 

 

 

 

 

 

 

 

 

3.In the Calibration menu select AUTOSCALE.

4.In the Auto-Measure menu verify that the waveform is approximately 0.8 Vp_p at approximately 1000 Hz. Record the reading in the performance test record.

3-3