HP 16532A - General Information
Table
EQUIPMENT |
| CRITICAL SPECIFICATIONS | RECOMMENDED MODEL | USE* | |
|
|
|
| ||
Signal Generator | Frequency: 1 - 250 MHz at approx . | HP 8656B Option 001 | P | ||
| 170 mV RMS resolution |
|
| ||
| Output Accuracy: ± 1 dB |
|
| ||
| 1 MHz time base accuracy 0.25 ppm |
|
| ||
|
|
|
| ||
Dc Power Supply | Range: − 35.000 to + 35.000 Vdc, ± 1 mV | HP 6114A | P | ||
|
|
|
|
| |
Digital Multimeter | 0.1 mV resolution |
| HP 3458A | P | |
| Accuracy: better than 0.005% |
|
| ||
| Resistance measurement: better than |
|
| ||
| 0.25% accuracy |
|
|
| |
|
|
|
| ||
Power Meter/Sensor | 1 - 250 MHz ± 3% accuracy | HP 436/8482A | P | ||
|
|
|
| ||
Power Splitter | Outputs differ by < 0.15 dB | HP 11667B | P | ||
|
|
|
|
| |
Blocking Capacitor | 50 | Ω BNC | HP 10240B | P | |
|
|
|
|
|
|
Adapter | 50 | Ω BNC (m)(m) |
| HP | C |
|
|
|
|
| |
Adapter | 50 | Ω Type | HP | P | |
|
|
|
|
| |
Adapter | 50 | Ω | HP | P | |
|
|
|
|
| |
Adapter | 50 | Ω Type | HP | P | |
|
|
|
|
| |
Tee Adapter | 50 | Ω BNC (m)(f)(f) | HP | P,C | |
|
|
|
| ||
Cable | Type | HP 11500B | P | ||
|
|
|
|
| |
Cable | 50 | Ω BNC | HP 10503A | P | |
|
|
|
|
| |
Cable | 50 | Ω BNC | HP 10502A | C | |
|
|
|
|
|
|
Cable | 50 | Ω |
| HP | C |
|
|
|
|
| |
*P= Performance Tests | C= Calibrations |
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|