HP 16532A - Performance Tests

3-13. TRIGGER SENSITIVITY

This test checks channel triggers for sensitivity at rated bandwidth.

Specification:

dc to 50 MHz : 0.063 x full scale (0.25 division)

50 to 250 MHz : 0.125 x full scale (0.5 division)

Equipment Required:

The following equipment is required for this test. Procedures are based on the model or part number rec- ommended.

 

 

 

Recommended

Equipment Required

 

Critical Specification

Model/Part Number

 

 

 

Signal Generator

50 and 250 MHz, 30-80 mV rms output

HP 8656B Opt. 001

Cable

50

Ω BNC 48-inch

HP 10503B

Adapter

50

Ω Type N (m) to BNC (f)

HP 1250-0780

 

 

 

 

 

 

 

 

Procedure:

1.In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. On the oscilloscope module set the following parameters in the order given:

 

Menu

 

Selection

 

Setting

 

 

 

 

 

 

Channel

 

Input

 

1

 

 

 

Probe

 

1:1

 

 

v/Div

 

400 mV

 

 

Offset

 

0 V

 

 

Coupling

 

50

Ω / DC

 

 

s/Div

 

20 ns

Trigger

 

Mode

 

Edge

 

 

Source

 

Chan 1

 

 

Level

 

0 V

Display

 

Mode

 

Average

 

 

Average #

 

8

 

 

 

Grid

 

On

 

 

 

Markers

 

Off

 

 

 

 

 

 

 

 

 

 

 

 

 

2. With an N-to-BNC adapter and BNC cable, connect the signal generator to channel 1 input.

3-15

Page 41