HP 16532A - Performance Tests
3.Select RUN - REPETITIVE. Wait approximately 15 seconds (averaging complete), then select STOP. The display should read 0.00 V ± 320.0 mV as read from the markers voltage field. Record the reading in the performance test record.
4.In the Channel menu, change the vertical sensitivity to 1 V/Div. Select RUN - REPETITIVE and wait
approximately 15 seconds (averaging complete), then select STOP. The display should read 0.00 V
± 80.0 mV. Record the reading in the performance test record.
5.In the Channel menu, change the vertical sensitivity to 100 mV/Div. Select RUN - REPETITIVE and
wait approximately 15 seconds (averaging complete), then select STOP. The display should read 0.00 V ± 8.0 mV. Record the reading in the performance test record.
6.In the Channel menu, change the vertical sensitivity to 10 mV/Div. Select RUN - REPETITIVE and
wait approximately 15 seconds (averaging complete), then select STOP. The display should read 0.00 V ± 800.0 μV. Record the reading in the performance test record.
DC Input Offset
Set the Channel Coupling field to 1MΩ / DC or damage to the equipment will result.
7.Use the
1input. Monitor the power supply using a voltmeter.
8.Use the following table for the next steps:
V/Div | Offset | Supply | Min | Max |
|
|
|
|
|
1 V/Div | − 35.00 V | − 35.00 V | − 35.4 V | − 34.6 V |
100 mV/Div | − 10.00 V | − 10.00 V | − 10.1 V | − 9.90 V |
20 mV/Div | − 2.00 V | − 2.00 V | − 2.02 V | − 1.98 V |
20 mV/Div | + 2.00 V | + 2.00 V | + 1.98 V | + 2.02 V |
100 mV/Div | + 10.00 V | + 10.00 V | + 9.90 V | + 10.1 V |
1 V/Div | + 35.00 V | + 35.00 V | + 34.6 V | + 35.4 V |
|
|
|
|
|
|
|
|
|
|
9.In the Channel menu set the V/Div range and offset per the first line of the table. Set the power supply per the first line of the table.
10.On the oscilloscope, select RUN - REPETITIVE and wait approximately 15 seconds (averaging complete), then STOP.
11.Read the voltage from the markers voltage field. It should be within the limits given in the table. Re- cord the reading in the performance test record.
12.Repeat steps 7 through 10 for the remaining lines of the table using the V/Div range, offset, and sup- ply voltages given in the table.
13.In the Channel menu select channel 2. In the Waveform Selection menu delete channel 1 and insert channel 2.
14.Repeat steps 1 through 12 for channel 2, replacing channel 1 with channel 2 where applicable.