5. On the HP 54502A Oscilloscope, set the following parameters, in the order given.

Menu

Selection

Setting

Channel

Probe

1.000:1

 

V/Div

200 mV

 

Offset

0.0 V

 

Coupling

DC

Timebase

s/Div

10 ns

 

Acquisition

Realtime

Trigger

Mode

Edge

 

Level

200 mV

 

Slope

Falling

 

 

 

 

 

 

6.In the HP 16532A Oscilloscope trigger menu, select Auto-Trig On. Verify that the 100 MHz System Clock is gated.

7.Select STOP on both the HP 54502A and HP 16532A oscilloscopes.

8.Disconnect the cable from both oscilloscopes.