HP 16532A - Performance Tests

3-11. BANDWIDTH

This test checks the bandwidth of the oscilloscope module.

Specification:

Bandwidth (dc coupled) dc to 250 MHz

Equipment Required:

The following equipment is required for this test. Procedures are based on the model or part number rec- ommended.

 

 

 

Recommended

Equipment Required

 

Critical Specification

Model/Part Number

 

 

 

 

Signal Generator

1

- 250 MHz at approx 170 mVrms

HP 8656B

Power Meter/Sensor

1

- 250 MHz ± 3% accuracy

HP 436/8482A

Power Splitter

Outputs differ by < 0.15 dB

HP 11667B

Cable

Type N (m to m) 24-inch

HP 11500B

Adapter

50 Ω Type N (m) to BNC (m)

HP 1250-0082

 

 

 

 

 

 

 

 

Procedure:

1.With the N cable, connect the signal generator to the power splitter input. Connect the power sensor to one output of the power splitter.

2.Using an N-to-BNC adapter, connect the other power splitter output to the Channel 1 input of the oscilloscope module.

3.In the Waveform Selection menu delete channel 2. If channel 1 is not inserted, insert channel 1. Set the following parameters.

 

Menu

 

Selection

 

Setting

 

 

 

 

 

 

Channel

 

Input

 

1

 

 

 

Probe

 

1:1

 

 

v/Div

 

80 mV

 

 

Offset

 

0 V

 

 

Coupling

 

50

Ω / DC

 

 

s/Div

 

200 ns

Trigger

 

Mode

 

Edge

 

 

Level

 

0 V

Display

 

Mode

 

Average

 

 

Average #

 

32

 

 

 

Grid

 

On

 

 

 

Markers

 

Off

 

 

 

 

 

 

 

 

 

 

 

 

 

3-11