Teledyne Model 9110TH NOx Analyzer Calibration Procedures, Calibration Quality Analysis

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 193
Image 193

Model 9110TH NOx Analyzer

Calibration Procedures

9. CALIBRATION PROCEDURES

This section contains information for calibrating a 9110T NOx Analyzer as well as other supporting information. For information on EPA protocol calibration, please refer to Section 10. This section is organized as follows:

SECTION 9.1 – Before Calibration

This section contains general information you should know before about calibrating the analyzer.

SECTION 9.2 – Manual Calibration Checks and Calibration of the 9110T Analyzer in its Base Configuration

This section describes:

The procedure for checking the calibrating of the 9110T and calibrating the instrument with no zero/span valves installed or if installed, not operating. It requires that zero air and span gas be installed through the Sample port.

Instructions for selecting the reporting range to be calibrated when the 9110T analyzer is set to operate in either the IND or AUTO reporting range modes.

SECTION 9.3 – Manual Calibration with the Internal Span Gas Generator

This section describes:

The procedure for manually checking the calibration of the instrument with optional internal span gas generator installed.

The procedure for manually calibrating the instrument using the optional internal span gas generator.

This practice is not approved by the US EPA.

SECTION 9.4 – Manual Calibration and Cal Checks with the Valve Options Installed This section describes:

The procedure for manually checking the calibration of the instrument with optional zero/span valves option installed.

The procedure for manually calibrating the instrument with zero/span valves and operating.

Instructions on activating the zero/span valves via the control in contact closures of the analyzers external digital I/O.

SECTION 9.5 – Automatic Zero/Span Cal/Check (AutoCal)

This section describes:

The procedure for using the AutoCal feature of the analyzer to check or calibrate the instrument.

The AutoCal feature requires that either the zero/span valve option or the internal span gas generator option be installed and operating.

SECTION 9.6 – Calibration Quality Analysis

Teledyne Analytical Instruments

193

Page 193
Image 193
Teledyne instruction manual Model 9110TH NOx Analyzer Calibration Procedures, Calibration Quality Analysis