Teledyne 9110T instruction manual Level 1 Calibrations Versus Level 2 Checks

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 224
Image 224

Model 9110T NOx Analyzer

EPA Protocol Calibration

Any repairs which might affect its calibration.

Physical relocation of the analyzer.

Any other indication (including excessive zero or span drift) of possible significant inaccuracy of the analyzer.

Following any of the activities listed above, the zero and span should be checked to determine if a calibration is necessary. If the analyzer zero and span drifts exceed the calibration limits in Section 12 of the Q.A. Handbook6, a calibration should be performed.

10.4. LEVEL 1 CALIBRATIONS VERSUS LEVEL 2 CHECKS

All monitoring instruments are subject to some drift and variation in internal parameters and cannot be expected to maintain accurate calibration over long periods of time the EPA requires a schedule of periodic checks of the analyzer’s calibration be implemented. Zero and span checks must be used to document that the data remains within required limits. These checks are also used in data reduction and system validation.

A Level 1 Span check is used to document that the 9110T is within control limits and must be conducted every 2 weeks. A Level 2 Span Check is to be conducted between the Level 1 Checks on a schedule to be determined by the user.

Table 10-2: Definition of Level 1 and Level 2 Zero and Span Checks

LEVEL 1 ZERO AND SPAN CALIBRATION

LEVEL 2 ZERO AND SPAN CHECK

A Level 1 zero and span calibration is a simplified,

A Level 2 zero and span check is an "unofficial" check of an

two-point analyzer calibration used when analyzer

analyzer's response. It may include dynamic checks made with

linearity does not need to be checked or verified.

uncertified test concentrations, artificial stimulation of the

(Sometimes when no adjustments are made to the

analyzer's detector, electronic or other types of checks of a

analyzer, the Level 1 calibration may be called a

portion of the analyzer, etc.

zero/span check, in which case it must not be

Level 2 zero and span checks are not to be used as a basis for

confused with a Level 2 zero/span check.) Since

analyzer zero or span adjustments, calibration updates, or

most analyzers have a reliably linear or near-

adjustment of ambient data. They are intended as quick,

linear output response with concentration, they

convenient checks to be used between zero and span

can be adequately calibrated with only two

calibrations to check for possible analyzer malfunction or

concentration standards (two-point concentration).

calibration drift. Whenever a Level 2 zero or span check

Furthermore, one of the standards may be zero

indicates a possible calibration problem, a Level 1 zero and

concentration, which is relatively easily obtained

span (or multipoint) calibration should be carried out before any

and need not be certified. Hence, only one

corrective action is taken.

certified concentration standard is needed for the

If a Level 2 zero and span check is to be used in the quality

two-point (Level 1) zero and span calibration.

control program, a "reference response" for the check should be

Although lacking the advantages of the multipoint

obtained immediately following a zero and span (or multipoint)

calibration, the two-point zero and span

calibration while the analyzer's calibration is accurately known.

calibration--because of its simplicity--can be (and

Subsequent Level 2 check responses should then be compared

should be) carried out much more frequently.

to the most recent reference response to determine if a change

Also, two-point calibrations are easily automated.

in response has occurred. For automatic Level 2 zero and span

Frequency checks or updating of the calibration

checks, the first scheduled check following the calibration

relationship with a two-point zero and span

should be used for the reference response. It should be kept in

calibration improves the quality of the monitoring

mind that any Level 2 check that involves only part of the

data by helping to keep the calibration relationship

analyzer's system cannot provide information about the portions

more closely matched to any changes (drifts) in

of the system not checked and therefore cannot be used as a

the analyzer response.

verification of the overall analyzer calibration.

 

In addition, an independent precision check between 0.08 and 0.10 ppm must be carried out at least once every two weeks. Table 10-1 summarizes the quality assurance activities for routine operations. A discussion of each activity appears in the following sections.

Teledyne Analytical Instruments

224

Page 224
Image 224
Teledyne 9110T Level 1 Calibrations Versus Level 2 Checks, Definition of Level 1 and Level 2 Zero and Span Checks