Teledyne 9110T instruction manual Short Calibration Checks, ZERO/SPAN Check Procedures

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 233
Image 233

Model 9110T NOx Analyzer

 

 

 

EPA Protocol Calibration

 

 

Table 10-3: Activity Matrix for Data Quality

 

 

 

 

 

 

 

 

 

 

 

 

 

Characteristic

Acceptance Limits

 

 

Frequency and Method

 

Action if Requirements

 

 

 

 

of Measurement

 

are not Met

 

 

 

 

 

 

 

 

 

Shelter

Mean temperature between 22°C

Check thermograph chart

 

Mark strip chart for the

 

 

temperature

and 28°C (72° and 82° F), daily

weekly for variations greater

 

affected time period.

 

 

 

fluctuations not greater than

±

°

than ± 2°C (4°F)

 

Repair or adjust

 

 

 

 

2 C

 

 

 

 

 

 

 

 

 

 

temperature control.

 

 

Sample

No moisture, foreign material,

 

Weekly visual inspection

 

Clean, repair, or replace as

 

 

introduction

leaks, obstructions; sample line

 

 

needed.

 

 

system

connected to manifold

 

 

 

 

 

 

 

Recorder

Adequate ink & paper

 

 

Weekly visual inspection

 

Replenish ink and paper

 

 

 

Legible ink traces

 

 

 

 

supply.

 

 

 

Correct chart speed and range

 

 

Adjust time to agree with

 

 

 

Correct time

 

 

 

 

clock; note on chart.

 

 

 

 

 

 

 

 

 

 

Analyzer

TEST measurements at nominal

Weekly visual inspection

 

Adjust or repair as needed.

 

 

operational

values

 

 

 

 

 

 

 

settings

9110T in SAMPLE mode

 

 

 

 

 

 

 

Analyzer

Zero and span within tolerance

Level 1 zero/span every

 

Find source of error and

 

 

operational

limits as described in Section 12 of

2 weeks; Level 2 between

 

repair.

 

 

check

the Q.A. Handbook6

 

 

Level 1 checks at frequency

 

After corrective action, re-

 

 

 

 

 

 

desired by user

 

calibrate analyzer.

 

 

Precision check

Assess precision as described in

Every 2 weeks, Subsection

 

Calc, report precision,

 

 

 

Sections 15&18 of the Q.A.

 

 

3.4.3 (Ibid.)

 

Section 12 of the Q.A.

 

 

 

Handbook6

 

 

 

 

Handbook6.

 

10.8.2. SHORT CALIBRATION CHECKS

A system of Level 1 and Level 2 zero/span checks (Table 10-2) is recommended. These checks must be conducted in accordance with the specific guidance given in Section 12 of the Q.A. Handbook6. Level 1 zero and span checks must be conducted every two weeks. Level 2 checks should be conducted in between the Level 1 checks at a frequency desired by the user. Span concentrations for both levels should be between 70 and 90% of the measurement range.

Zero and span data are to be used to:

Provide data to allow analyzer adjustment for zero and span drift;

Provide a decision point on when to calibrate the analyzer;

Provide a decision point on invalidation of monitoring data.

These items are described in detail in Sections 15 & 18 of the Q.A. Handbook6. Refer to Section 11 of this manual if the instrument is not within the allowed margins. We recommend using APICOM and the DAS for analysis and documentation of zero/span check data.

10.8.3. ZERO/SPAN CHECK PROCEDURES

The Zero and span calibration can be checked in a variety of different ways. They include:

Manual zero/span checks can be done from the front panel touchscreen. The procedure is in Section 9.2.2 of this manual.

Automatic zero/span checks can be performed every night. See Section 9.5 of this manual for setup and operation procedures.