Model 9110T NOx Analyzer

Data Acquisition System (DAS) and APICOM

Note

The DAS can be disabled only by disabling or deleting its individual data

 

channels.

7.1. DAS STRUCTURE

The DAS is designed around the feature of a “record”. A record is a single data point. The type of data captured in a record are defined by two properties:

PARAMETER type that defines the kind of data to be stored (e.g. the average of O3 concentrations measured with three digits of precision). See Section 7.1.3.3.

A TRIGGER event that defines when the record is made (e.g. timer; every time a calibration is performed, etc.). See Section 7.1.3.2.

The specific PARAMETERS and TRIGGER events that describe an individual record are defined in a construct called a DATA CHANNEL (see Section 7.1.3). Each data channel relates one or more parameters with a specific trigger event and various other operational characteristics related to the records being made (e.g. the channels name, number or records to be made, time period between records, whether or not the record is exported via the analyzer’s RS-232 port, etc.).

7.1.1. DAS CHANNELS

The key to the flexibility of the DAS is its ability to store a large number of combinations of triggering events and data parameters in the form of data channels. Users may create up to 20 data channels and each channel can contain one or more parameters. For each channel, the following are selected:

one triggering event

up to 50 data parameters, which can be the shared between channels.

several other properties that define the structure of the channel and allow the user to make operational decisions regarding the channel.

Table 7-2: DAS Data Channel Properties

PROPERTY

DESCRIPTION

DEFAULT

SETTING RANGE

SETTING

 

 

 

NAME

The name of the data channel.

“NONE”

Up to 6 letters or digits1.

TRIGGERING

The event that triggers the data channel to

ATIMER

Any available event

EVENT

measure and store the datum.

(see Appendix A-5).

 

NUMBER AND

A user-configurable list of data types to be

1

Any available parameter

LIST OF

recorded in any given channel.

(PMTDET)

(see Appendix A-5).

PARAMETERS

 

 

 

 

The amount of time between each channel data

000:01:00

000:00:01 to

REPORT PERIOD

366:23:59

point.

(1 hour)

 

(Days:Hours:Minutes)

 

 

 

NUMBER OF

The number of reports that will be stored in the

100

1 to 1 million, limited by

data file. Once the limit is exceeded, the oldest

RECORDS

available storage space.

data is over-written.

 

 

 

 

RS-232 REPORT

Enables the analyzer to automatically report

OFF

OFF or ON

channel values to the RS-232 ports.

 

 

 

Teledyne Analytical Instruments

164

Page 164
Image 164
Teledyne 9110T instruction manual DAS Structure, DAS Channels