Teledyne 9110T instruction manual Pneumatic Layout for Zero Scrubber and IZS OPT 50G

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 71
Image 71

Model 9110TH NOx Analyzer

Getting Started

HNO3 is a liquid at room temperature, so once the HNO3 is released by the permeation tube it condenses and collects along the 9110T’s wetted surfaces, While liquid HNO3 does not directly effect the quality of NOx measurements of the Model 9110T, it does give off small amounts of gaseous HNO3 which is converted into NO by the 9110T’s NOx NO converter resulting in an artificially high NO2 concentration by 8% to 12%. This is particularly bothersome when 9110T is attempting to measure a zero point, such as during calibration, since the NO2 concentration will only reach a true zero point once the majority of the HNO3 coating the wetted surfaces has reverted to NO2 and this can take a very long time.

The 9110T includes a special HNO3 desorbed which eliminates any HNO3 given off by the permeation tube before it can be converted into NO by the analyzer’s converter.

PNEUMATIC LAYOUT FOR ZERO SCRUBBER AND IZS (OPT 50G)

Figure 3-27: Pneumatics with the Internal Span Gas Generator (OPT 50G)

Table 3-11: Internal Span Gas Generator Valve Operating States OPT 50G

MODE

VALVE

CONDITION

VALVE PORT

STATUS

 

 

 

SAMPLE

Sample/Cal

Open to SAMPLE inlet

3 2

Teledyne Analytical Instruments

71

Page 71
Image 71
Teledyne 9110T instruction manual Pneumatic Layout for Zero Scrubber and IZS OPT 50G