Model 9110TH NOx Analyzer

Setup Menu

These parameters can also be captured in the internal Data Acquisition System (DAS); refer to Appendix A for Analog-In DAS parameters.

To adjust settings for the Analog Inputs option parameters press:

DIAG

ANALOG I / O CONFIGURATION

 

 

 

PREV

NEXT

 

ENTR

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

AOUTS CALIBRATED: NO

 

 

 

 

 

 

 

 

 

Press SET> to scroll to the first

< SET SET>

CAL

EXIT

 

channel. Continue pressing SET>

 

 

 

 

 

 

to view each of 8 channels.

 

 

 

 

 

 

 

 

 

 

 

 

DIAG AIO

XIN1:1.00,0.00,V,OFF

 

 

 

 

< SET SET>

EDIT

EXIT

 

 

Press EDIT at any channel

 

 

to to change Gain, Offset,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Units and whether to display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the channel in the Test

 

 

 

 

 

 

 

functions (OFF/ON).

 

 

 

DIAG AIO

XIN1 GAIN:1.00V/V

 

 

 

 

 

SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

DIAG AIO

XIN1 OFFSET:0.00V

 

< SET

SET>

EDIT

EXIT

 

 

 

 

 

 

DIAG AIO

XIN1 UNITS:V

 

< SET

SET>

EDIT

EXIT

 

 

 

 

 

 

DIAG AIO

XIN1 DISPLAY:OFF

 

< SET

 

EDIT

EXIT

 

 

 

 

DIAG AIO

 

 

XIN1 GAIN:1.00V/V

+ 0

0

1

.0

0

ENTR EXIT

 

 

 

 

 

 

Press to change

Gain value

Pressing ENTR records the new setting

and returns to the previous menu.

Pressing EXIT ignores the new setting and

returns to the previous menu.