Model 9110T NOx Analyzer

Troubleshooting & Service

5. Alternately, turn on and off the output noting the voltage on the voltmeter.

It should vary between 0 volts for ON and 5 volts for OFF.

Table 12-10: Status Outputs Check

PIN (LEFT TO RIGHT)

STATUS

1

ST_SYSTEM_OK

2

ST_CONC_VALID

3

ST_HIGH_RANGE

4

ST_ZERO_CAL

5

ST_SPAN_CAL

6

ST_DIAG_MODE

7

Not Used on 9110T

8

ST_O2_CAL

 

 

12.7.6.4. Control Inputs

The control input bits can be tested by applying a trigger voltage to an input and watching changes in the status of the associated function under the SIGNAL I/O submenu:

EXAMPLE: to test the “A” control input:

1.Under the DIAGSignal I/O menu (see Section 12.1.3), scroll through the inputs and outputs until you get to the output named EXT_ZERO_CAL.

2.Connect a jumper from the “+” pin on the appropriate connector to the “U” on the same connector.

3.Connect a second jumper from the “” pin on the connector to the “A” pin.

4.The status of EXT_ZERO_CAL should change to read “ON”.

5.Connect a second jumper from the “” pin on the connector to the “B” pin.

6.The status of EXT_ZERO_CAL should change to read “ON”.

Table 12-11: 9110T Control Input Pin Assignments and Corresponding Signal I/O Functions

INPUT

CORRESPONDING I/O SIGNAL

A

EXT_ZERO_CAL

B

EXT_SPAN_CAL1

C, D, E& F

NOT USED