Model 9110T NOx Analyzer

Getting Started

3.4.STARTUP, FUNCTIONAL CHECKS, AND INITIAL CALIBRATION

CAUTION!

If the presence of ozone is detected at any time, power down the instrument

and contact TAI Customer Service as soon as possible:

(626) 934-1500 or email: ask_TAI@teledyne.com

If you are unfamiliar with the 9110T principles of operation, we recommend that you read Section 13. For information on navigating the analyzer’s software menus, see the menu trees described in Appendix A.

3.4.1. START UP

After the electrical and pneumatic connections are made, an initial functional check is in order. Turn on the instrument. The pump and exhaust fan should start immediately. The display will show a splash screen and other information during the initialization process while the CPU loads the operating system, the firmware and the configuration data.

The analyzer should automatically switch to Sample Mode after completing the boot-up sequence and start monitoring the gas. However, there is an approximately one hour warm-up period before reliable gas measurements can be taken. During the warm-up period, the front panel display may show messages in the parameters (Param) field.