Model 9110T NOx Analyzer

Getting Started

Figure 3-7: Analog In Connector

Pin assignments for the Analog In connector are presented in Table 3-4.

Table 3-4: Analog Input Pin Assignments

PIN

DESCRIPTION

DAS

PARAMETER1

1

Analog input # 1

AIN 1

 

 

 

2

Analog input # 2

AIN 2

3

Analog input # 3

AIN 3

4

Analog input # 4

AIN 4

5

Analog input # 5

AIN 5

 

 

 

6

Analog input # 6

AIN 6

 

 

 

7

Analog input # 7

AIN 7

8

Analog input # 8

AIN 8

GND

Analog input Ground

N/A

1 See Section 7 for details on setting up the DAS.

3.3.1.3. Connecting Analog Outputs

The 9110T is equipped with several analog output channels accessible through a connector on the back panel of the instrument.

Output channels A1, A2 and A3 are assigned to the NOx, NO and NO2 concentration signals of the analyzer.

The default analog output voltage setting of these channels is 0 to 5 VDC with a reporting range of 0 to 500 ppb.

An optional Current Loop output is available for each.

The output labeled A4 is special. It can be set by the user to output any one a variety of diagnostic test functions (see Section 5.9.4).

The default analog output voltage setting of these channels is also 0 to 5 VDC.

See Section 5.9.4 for a list of available functions and their associated reporting range.

There is no optional Current Loop output available for Channel A4.

To access these signals attach a strip chart recorder and/or data-logger to the appropriate analog output connections on the rear panel of the analyzer. Pin-outs for the analog output connector are:

Teledyne Analytical Instruments

42