Model 9110T NOx Analyzer

Instrument Maintenance

CAUTION!

The following procedures apply only to the External Zero Air Scrubber and NOT to the inline exhaust scrubber cartridge that is part of the pump pack assembly.

IMPORTANT

IMPACT ON READINGS OR DATA

This procedure can be carried out while the instrument is running,

however ensure that the analyzer is not in ZERO calibration mode.

1.

Locate the scrubber on the outside rear panel; Figure 11-4 shows the exploded

 

assembly.

2.

Remove the old scrubber by disconnecting the 1/4” plastic tubing from the DFU

 

particle filter using 9/16” and 1/2" wrenches.

3.

Remove the DFU particle filter from the cartridge using 9/16” wrenches.

4.

Unscrew the top of the scrubber canister and discard the Purafil© and charcoal

 

contents.

 

Ensure to abide to local laws about discarding these chemicals.

 

The rebuild kit (listed in Appendix B) comes with a Material and Safety Data

 

Sheet, which contains more information on these chemicals.

5.

It is not necessary to remove the insert from the barrel, but if removed, perform the

 

following procedure:

 

Coat the threads of the insert with epoxy (TAI P/N CH32).

 

Hand tighten insert to barrel.

6.

It is not necessary to remove the nylon tube fitting from the insert, but if removed,

 

apply Teflon tape (TAI P/N HW36) to the threads of the nylon tube fitting before

 

installing on the insert.

7.

Refill the scrubber with charcoal at the bottom and the Purafil© chemical at the top.

 

Use three, white retainer pads to separate the chemicals as shown Figure 11-4

8.

Replace the screw-top cap and tighten the cap; hand-tighten only.

9.

If necessary, replace the filter with a new unit and discard the old. See Section

 

11.3.7.1.

 

The bottom retainer pad should catch most of the dust, the filter should not be

 

visibly dirty (on the inside).

10.

Replace the scrubber assembly into its clips on the rear panel.

11.

Reconnect the plastic tubing to the fitting of the DFU particle filter.

12.

Adjust the scrubber cartridge such that it does not protrude above or below the

 

analyzer in case the instrument is mounted in a rack.

Teledyne Analytical Instruments

250

Page 250
Image 250
Teledyne 9110T instruction manual Teledyne Analytical Instruments 250