Teledyne 9110T instruction manual Evaluating NO2  no Converter Performance, Step

Models: 9110T

1 378
Download 378 pages 16.38 Kb
Page 295
Image 295

Model 9110T NOx Analyzer

Troubleshooting & Service

STEP 3:

To cause the analyzer to calculate and record the NO2 NO converter efficiency, press:

Starting from

CONVERTER EFFICIENCY MENU

(see preceding steps)

Toggle these

buttons to initialize

the converter

efficiency at 1.0000.

M-P CAL CONVERTER EFFICIENCY MENU

 

 

NO2

CAL

SET

 

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

CE FACTOR:1000.0 Gain

 

 

 

1.

0

0

0

0

 

ENTR EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

CONVERTER EFFICIENCY MENU

 

 

NO2

CAL

SET

 

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

 

RANGE=500.0 PPB

NOX= XXXX

 

 

< TST TST >

ENTR

 

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The analyzer calculates the converter’s efficiency.

Check the calculated efficiency

gain factor.

Toggle TST> button until ...

SAMPLE

NO2 STB=XX.X PPB

<TST TST>

 

SETUP

 

 

 

 

 

 

Allow NO2 gas of the proper concetration to enter

the sample port at the rear of the analyzer.

 

SAMPLE

 

NO2 STB=XX.X PPB

 

 

 

<TST TST> ENTR

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

CONVERTER EFICIENCY MENU

 

 

NO2 CAL

SET

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set the Display to show

the NOX STB test

function.

This function calculates

the stability of the NO/NOx

measurement.

Wait until NOX STB falls below 0.5 ppb.

This may take several

minutes.

If the gain factor is NOT

 

 

 

 

 

 

M-P CAL

 

CE FACTOR=0.9852 Gain

between 0.9600 and 1.0200,

 

 

 

 

 

0.

8

8

5

2

ENTR EXIT

 

the

NO2 NO converter

 

 

 

 

 

 

needs to be replaced.

 

 

 

 

 

 

12.7.10.4. Evaluating NO2 NO Converter Performance

If the converter appears to have performance problems (conversion efficiency is outside of allowed range of 96-102%), check the following:

Teledyne Analytical Instruments

295

Page 295