Model 9110T NOx Analyzer

Principles of Operation

13.7.1.1. AC Configuration – Internal Pump (JP7)

If your 9110T includes an internal pump the following table, jumper set JP7 is used to configure the power supplied to it as shown in Figure 13-27.

Table 13-6: AC Power Configuration for Internal Pumps (JP7)

LINE

LINE

JUMPER

 

JUMPER

FUNCTION

BETWEEN

POWER

FREQUENCY

COLOR

 

PINS

 

 

 

 

 

 

 

 

 

 

 

 

Connects pump pin 3 to 110 / 115 VAC power line

2 to 7

 

60 HZ

WHITE

 

 

 

Connects pump pin 3 to 110 / 115 VAC power line

3 to 8

110VAC

 

 

Connects pump pins 2 & 4 to Neutral

4 to 9

115 VAC

 

 

Connects pump pin 3 to 110 / 115 VAC power line

2 to 7

 

50 HZ1

BLACK

 

 

 

Connects pump pin 3 to 110 / 115 VAC power line

3 to 8

 

 

 

Connects pump pins 2 & 4 to Neutral

4 to 9

 

 

 

 

 

 

60 HZ

BROWN

Connects pump pins 3 and 4 together

1 to 6

220VAC

Connects pump pin 1 to 220 / 240VAC power line

3 to 8

 

 

240 VAC

50 HZ1

BLUE

Connects pump pins 3 and 4 together

1 to 6

 

 

 

 

Connects pump pin 1 to 220 / 240VAC power line

3 to 8

 

 

 

 

 

 

 

 

1 A jumper between pins 5 and 10 may be present on the jumper plug assembly, but has no function on the Model 9110T.

110 VAC /115 VAC

 

220 VAC /240 VAC

1

6

 

1

6

2

7

 

2

7

3

 

8

3

8

4

9

 

4

9

5

10

 

5

10

May be present on 50 Hz version of jumper set, but is not functional on the T200

Figure 13-27: Pump AC Power Jumpers (JP7)

Teledyne Analytical Instruments

360

Page 360
Image 360
Teledyne 9110T instruction manual AC Power Configuration for Internal Pumps JP7